| CRM-510LP Specifications |
| Radiation Detected: |
Radon-222 Alpha |
| Detector: |
Air Ionization Probe |
| Dynamic Range: |
0.5 to 2000 pCi/l (18 to 74000 Bq/m3) |
| Units: |
pCi/l or Bq/m3 |
| Sensitivity: |
0.3 CPM Per pCi/l (Nominal) (0.008 CPM Per Bq/m3) |
| Linearity: |
0.5 to 2000 pCi/l (<10% Deviation) |
| Sampling Mode: |
Passive Air Diffusion |
| Operating Modes: |
|
| Data Collection |
Elapse Time Displayed |
| Data Printing |
ASCII text Data, Seiko/Epson® Graphics Data |
| Data Storage: |
192 One Hour Data Points (8 Days) |
| |
|
| Environmental Sensors: |
|
| Temperature |
Range 50-104 ºF (10-40 ºC) Sensitivity ± 1 ºF (± 0.5 ºC) |
| Barometric Pressure |
Range 0-38 "Hg (0-127 kPa) Sensitivity ± 0.2 "Hg (± 0.5 kPa) |
| Relative Humidity |
Range 10-90% Sensitivity ± 5% |
| |
|
| Display: |
|
| Alphanumeric Prompts |
16-Digit one Line LCD |
| Low Battery Indicator |
LCD Message |
| Pulse Indicator |
Red LCD |
| Controls: |
|
| Key Switch |
Off, I/O, and Run Positions |
| Two Momentary Push Buttons |
Multifunction Select & Execute |
| External Output (Male DB-25P) |
|
| Serial |
RS-232C |
| Parallel |
8 Bit Simplified Centronics |
| Power: |
|
| Electrometer |
Four 9 V Alkaline (Nominal Life 1 Yr.) |
| Computer/Data Logger |
7.2 V, 7.2 Ah Lithium (Nominal Life 1 Yr.) |
| Size: |
|
| Height (w/hangle) |
5.4" (137 mm) |
| Width |
6.6" (160 mm) |
| Depth |
7.4" (190 mm) |
| Weight |
3.9 pounds (1.8 Kg) |
| Environmental Operating Range |
50-104 ºF (10-40 ºC) |
| Noncondensing |
10 to 90% RH noncondensing |